Abstract Body (Do not enter title and authors here): Background
Tricuspid regurgitation (TR) can lead to right ventricular failure (RVF) and poor clinical outcomes. The TriClip™ G4 transcatheter edge-to-edge repair (TEER) system has shown improved outcomes in TR patients and received FDA approval on April 2, 2024. The FDA’s Manufacturer and User Facility Device Experience (MAUDE) database collects mandatory manufacturer reports and voluntary provider reports on device-related complications.
Methods
We retrieved all TriClip™-related adverse events from the MAUDE database post-FDA approval, excluding duplicate entries and literature reviews. After removing 47 literature reviews and 36 duplicate reports, 443 unique adverse event reports from 407 procedures (some with multiple events) were analyzed between April 2, 2024, and January 31, 2025.
Results
SLDA (Single Leaflet Device Attachment) was the most frequent complication, occurring in 204 cases (Figure 1). Detached leaflets included septal (105 cases), anterior (46), posterior (27), lateral (3), multiple (2), and unreported. Contributing factors included poor imaging, tethered leaflet, clip-clip interaction, dense chordae, large right atrium (RA) or gap size, pacer lead, or prior valve intervention. Persistent TR was a common consequence. SLDA was addressed with additional clips in 77 cases (18 requiring multiple clips), no intervention in 55 cases (13 with significant residual TR), repeat TriClip procedures in 16 cases, and surgical repair in 4 cases. Other complications included partial SLDA (23 cases), clip deployment or gripper malfunctions (68), chordal entanglement (41), leaflet entanglement (3), and clip migration/embolization (6). Migrated clips were found in the RA, right atrial appendage, cardiac vein, or pulmonary artery; three required snare retrieval, and three were left in place. Triclip Steerable Guide Catheter (TSGC) malfunctions, such as leaks, loss of column support, and blockages, were reported in 29 cases, with 27 requiring replacement. Miscellaneous events included residual TR, leaflet grasping issues, F-knob malfunction, lock line jams, and TSGC entanglement. Clip opening during Establishing Final Arm Angle (EFAA) occurred in 25 cases, and leaflet puncture in 15 cases, with one requiring surgical intervention.
Conclusion
SLDA was the most frequently reported complication following TriClip™ approval, with some cases leading to severe outcomes. Continued surveillance is crucial to assess long-term safety and optimize procedural strategies.
  • Gupta, Lovish  ( Allegheny General Hospital , Pittsburgh , Pennsylvania , United States )
  • Gulati, Arvind  ( Mayo Clinic , Rochester , Minnesota , United States )
  • Futela, Pragyat  ( MetroHealth , Cleveland , Ohio , United States )
  • Khanna, Tejasvini  ( Corewell Health , Royal Oak , Michigan , United States )
  • Aneja, Ashish  ( MetroHealth , Cleveland , Ohio , United States )
  • Kowlgi, Gurukripa  ( Mayo Clinic , Rochester , Minnesota , United States )
  • Gupta, Tanush  ( University of Vermont Medical Center , Burlington , Vermont , United States )
  • Chatterjee, Arka  ( BUMC Tucson , Tucson , Arizona , United States )
  • Author Disclosures:
    Lovish Gupta: DO NOT have relevant financial relationships | Arvind Gulati: DO NOT have relevant financial relationships | Pragyat Futela: DO NOT have relevant financial relationships | Tejasvini Khanna: No Answer | Ashish Aneja: No Answer | Gurukripa Kowlgi: No Answer | Tanush Gupta: No Answer | Arka Chatterjee: No Answer
Meeting Info:

Scientific Sessions 2025

2025

New Orleans, Louisiana

Session Info:

Frontiers in Cardiac Care: Innovations, Inflammation, and Rare Disease Challenges

Monday, 11/10/2025 , 01:00PM - 02:00PM

Abstract Poster Board Session

More abstracts on this topic:
Evolution of Tricuspid Regurgitation and Clinical Outcomes in Patients with Heart Failure with Preserved Ejection Fraction: Insights from the PURSUIT-HFpEF Registry

Okada Masato, Nakatani Daisaku, Sotomi Yohei, Sakata Yasushi, Tanaka Nobuaki, Iwakura Katsuomi, Seo Masahiro, Hayashi Takaharu, Yano Masamichi, Yamada Takahisa, Yoshio Yasumura, Hikoso Shungo

Chronic Kidney Disease is Associated With Decreased Survival In Patients with Cardiogenic Shock and Tricuspid Regurgitation

Ramchandani Sunil, Oren Daniel, Nair Ranjit, Szinte Julia, Lorenzana Adrian, Pink Kevin, Zepeda Ignacio

More abstracts from these authors:
Predictors of Left Ventricular Function Recovery After Premature Ventricular Complex Ablation: A Retrospective Study

Futela Pragyat, Poddar Aastha, Gulati Arvind, Praveen Niharika, Ferreira Felix Iuri, Quelal Karol, Kowlgi Gurukripa

Early Reported Events with the TriClip™ System for Transcatheter Tricuspid Valve Repair: Insights from FDA's MAUDE Database

Khanna Tejasvini, Futela Pragyat, Bansal Kannu, Chatterjee Arka

You have to be authorized to contact abstract author. Please, Login
Not Available