The Ultimate Test in Hemocompatibility: HeartMate3 Restart After Prolonged Pump Shut Down

Abstract Body (Do not enter title and authors here): Introduction:
The HeartMate 3 (HM3) LVAD, was shown to have a higher survival free of hemocompatibility related adverse events (HRAE) compared to its predecessors (HM2, HVAD). Superior HM3 outcomes are attributed to wide blood flow pathways coupled with frictionless movement and intrinsic pulsatility, reducing shear stress and blood stasis. It is unknown if improved hemocompatibility can withstand pump restart after prolonged shutdown. We herein report a case of HM3 pump stoppage without subsequent HRAE.
Case Presentation:
A 41-year-old male underwent HVAD implant in 2019 for advanced non-ischemic cardiomyopathy. This was exchanged to HM3 for recurrent neurological events despite therapeutic anticoagulation. Ten months after exchange, he awakened one morning to find his LVAD had been off for an unknown period but had not heard any device alarms (85dB if on battery power, 165dB if on wall power). Since he felt well he changed to battery power resulting in immediate pump restart. Log file analysis showed pump stoppage 2 am – 10 am, without preceding low flow alarms or power elevations.
Management and Outcomes:
In the ER INR was 1.5 (2.8 one week prior) and systemic heparin was started. Evaluation included: 1) CT brain without acute infarcts 2) echocardiogram without intracardiac thrombus 3) CT Angiography with patency of the inflow cannula and outflow graft 4) stable serial LDH measurements. The controller was exchanged, and analysis noted normal function. 1-year later the patient is maintained on warfarin and aspirin 325mg without further HRAE. Given the patient’s neurologic history and pump stoppage event, we did not invoke ARIES trial guidance and thus continued aspirin.
Conclusion:
Pump stoppage occurs when there is complete battery depletion, disconnection of both power leads, or the percutaneous lead from system controller. Our case is unique in that the duration of pump shutdown was 8 hours and INR subtherapeutic, without HRAE in the background of neurologic events on HVAD support. It has been previously reported that complete outflow graft thrombosis can occur shortly after LVAD decommissioning. The HM3 User Manual recommends restarting the pump immediately if off for a few minutes and using clinical judgment for longer durations due to increased risk for thromboembolic events. Our case adds to the paucity of existing data on improved hemocompatibility of the HM3 during rare circumstances of the ultimate test in hemocompatibility: complete pump shut down.
  • Goral, Montana  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Patel, Ravi  ( Methodist Health System Dallas , Richardson , Texas , United States )
  • O'hara, Patrick  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Jawaid, Anas  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Hussain, Fizza  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Peltz, Matthias  ( UT SOUTHWESTERN MEDICAL CENTER , Dallas , Texas , United States )
  • Farr, Maryjane  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Araj, Faris  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Hightower, Mary  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Edwards, Michele  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Varghese, Densey  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Thomas, Melanie  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Hardin, Elizabeth  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Kim, Han Wool  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Malensek, Paris  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Beaini, Hadi  ( University of Texas Southwestern , Dallas , Texas , United States )
  • Author Disclosures:
    Montana Goral: DO NOT have relevant financial relationships | Ravi Patel: DO NOT have relevant financial relationships | Patrick O'Hara: DO NOT have relevant financial relationships | Anas Jawaid: DO NOT have relevant financial relationships | Fizza Hussain: No Answer | Matthias Peltz: DO NOT have relevant financial relationships | Maryjane Farr: DO have relevant financial relationships ; Consultant:TransMedics:Active (exists now) ; Consultant:Natera:Active (exists now) | Faris Araj: DO NOT have relevant financial relationships | Mary Hightower: DO NOT have relevant financial relationships | Michele Edwards: No Answer | densey varghese: No Answer | Melanie Thomas: DO NOT have relevant financial relationships | elizabeth hardin: No Answer | Han Wool Kim: DO NOT have relevant financial relationships | Paris Malensek: DO NOT have relevant financial relationships | Hadi Beaini: DO NOT have relevant financial relationships
Meeting Info:

Scientific Sessions 2024

2024

Chicago, Illinois

Session Info:

Crazy Clinical Cases in Heart Failure

Saturday, 11/16/2024 , 10:30AM - 11:30AM

Abstract Poster Session

More abstracts on this topic:
Association between small pre-implant left ventricular end diastolic diameter and post left ventricular assist device implantation all-cause mortality: A systematic review and meta-analysis

Pingili Adhvithi, Feller Erika, Vadiyala Mounika Reddy, Anmol Kumar, Jain Muskan, Varghese Jeffy Susan, Ahluwalia Vibhor, Prajapati Kesar, Chalasani Vardhan, Desai Rupak

Active Decompression during Automated Head-up Cardiopulmonary Resuscitation

Pourzand Pouria, Metzger Anja, Moore Johanna, Suresh Mithun, Salverda Bayert, Hai Hamza, Kaizer Alexander, Duval Sue, Bachista Kerry, Lurie Keith

More abstracts from these authors:
High Output Heart Failure After Leadless Pacemaker Placement In a Heart Transplant Recipient

Kim Han Wool, Araj Faris, Malensek Paris, Morlend Robert, Beaini Hadi, Patel Ravi, O'hara Patrick, Goral Montana, Hussain Fizza, Farr Maryjane

Demonstrating Pulmonary Hypertension Reversibility Before Heart Transplantation: A Novel Bedside Technique Using Postural Change

Patel Ravi, Araj Faris, Goral Montana, Hardin Elizabeth, Kim Han Wool, Malensek Paris, Beaini Hadi, O'hara Patrick, Jawaid Anas, Hussain Fizza

You have to be authorized to contact abstract author. Please, Login
Not Available